Calibration of instruments and pipettes
ISO 8655 Certificate
Menu
- Biological exterior decontamination (bactericide, virucide, fungicide and DNA / RNA)
- Initial inspection, tightness check.
- Calibration according to manufacturer following the method described in ISO 8655.
- Test in three different volumes performing 10 measurements. Multichannel pipettes are verified on all channels. Pipettes with out-of-tolerance results are subjected to an adjustment of the aspiration and ejection device following the manufacturer's protocol according to brand.
- After the adjustment is subjected to a new calibration.
- Delivery in airtight envelope with calibration certificate.

Working conditions

- Relative humidity: 40% - 60%.
- Ambient temperature: 19.5 - 24 ºC.
- Atmospheric pressure: 950 - 1050 mbar
- For all types of calibrations.
- Calibration and Repair is performed with original spare parts of all brands.
- Withdrawal and delivery in hand without charge.
- Pipettes are removed, with prior notice, in 24-48 hours and delivered within a maximum period of 5-7 days.